Attogene’s Universal Lateral Flow Assay Kit (Particles Not Included, BYOP, AU2064) is a lateral flow assay kit without particles, designed for developers who want to test their own streptavidin-conjugated reporter. Each kit contains 50 pre-striped 4.5 mm dipsticks for detecting dual-labeled DNA and RNA, including PCR, LAMP and RPA products, with results generally in 10–15 minutes.
Because you supply the reporter, the same strips can be used to compare colloidal gold or silver, europium beads, phycoerythrin, quantum dots, latex beads and other streptavidin-conjugated particles. There is no need to stripe capture antibodies and no expensive equipment is required.
Key Features
- Bring your own particles: compatible with user-supplied streptavidin reporter-particle conjugates, colorimetric or fluorescent.
- Pre-striped capture lines: anti-FITC/FAM, anti-DIG and anti-DNP test lines, depending on format, with a biotin control line.
- Multiplex up to 3 targets: the 3-line format detects two targets and the 4-line format detects three.
- Fast readout: reaction generally complete in 10–15 minutes; read by eye, by photograph or with a lateral flow reader.
- Cost-effective screening: evaluate reporters and labeling strategies before downstream lateral flow assay development.
Choose Your Format
- AU2064-01, 2 Line: test line anti-FITC/FAM; biotin control line. For biotin + FITC/FAM labeled products.
- AU2064-02, 3 Line: test line 1 anti-FITC/FAM, test line 2 anti-DIG; biotin control line. Multiplex up to two targets.
- AU2064-03, 4 Line: test line 1 anti-DNP, test line 2 anti-FITC/FAM, test line 3 anti-DIG; biotin control line. Multiplex up to three targets.
Inquire about custom configurations at sales@attogene.com.
What’s in the Kit
Each lateral flow assay kit without particles includes:
- 50 lateral flow dipsticks (4.5 mm) in a desiccated vial
- 20 µL control nucleic acid containing biotin and FAM (all formats)
- 20 µL control nucleic acid containing biotin and DIG (AU2064-02 and -03)
- 20 µL control nucleic acid containing biotin and DNP (AU2064-03)
- One 96-well plate
- Manual
You will also need: your streptavidin reporter-particle conjugate, a running buffer suited to your particles, a timer, a centrifuge for clearing specimens, and a pipettor with tips. Attogene offers a range of running buffers, including RNase-free variants, for colloidal gold, quantum dots, phycoerythrin and europium beads; AU2035 provides samples of five RNase-free running buffers for screening.
Who It’s For
This lateral flow assay kit without particles is for assay developers and researchers who already have, or want to evaluate, a specific reporter particle. Typical uses include comparing gold against fluorescent labels for sensitivity, qualifying a new conjugate lot, and testing a reporter with your own labeled amplification products before committing to a custom strip.
How to Use the Lateral Flow Assay Kit Without Particles
- Add 100–150 µL of running buffer to a well of the 96-well plate.
- Add your streptavidin reporter-particle conjugate. For a 10 OD colloidal gold solution, 2–10 µL is a typical starting point; for fluorescent particles at 1 mg/mL or 1 µM, start with 0.5–2 µL.
- Add your sample (1–5 µL is a good starting point; keep below 15 µL if possible). Run a positive control with 5 µL of control oligonucleotide and a blank negative control.
- Mix each well completely by pipetting up and down several times.
- Add one dipstick per well with the arrows facing up.
- Incubate for 10–15 minutes, then read by eye, by photograph or with a lateral flow reader.
See the AU2064 kit manual (PDF) for full details.
How It Works and Reading Results
Your streptavidin-conjugated particle binds the biotin label on the nucleic acid. Each target must also carry a capture label (FITC/FAM, DIG or DNP), so dual-labeled product is captured at its test line. Particles not bound to labeled product are captured at the biotin control line.
- Test line: intensity increases with the amount of labeled nucleic acid, compared with a buffer-only strip.
- Control line: intensity decreases as the test line increases, which is expected.
- Weak or no test line at high input: possible hook effect. A serial dilution can bring the product back into the working ratio with the particles.
The ratio of labeled nucleic acid to reporter particles is central to optimization, and the right amount depends on the purity and sequence of your product. A standard curve helps define the working range.
Frequently Asked Questions
Which particles can I use?
Any streptavidin-conjugated reporter, such as colloidal gold or silver, europium beads, phycoerythrin, quantum dots or latex beads.
Is running buffer included?
The manual for this lateral flow assay kit without particles lists dipsticks, controls, a 96-well plate and the manual. Buffers matched to your particle type are available separately; contact sales@attogene.com for options.
What labels does my product need?
Biotin plus FITC/FAM, DIG or DNP on the same molecule, depending on the format you choose.
How do I store the kit?
Store at 2–4 °C until use and keep the strips in the desiccated vial until you are ready to test.
Is there a version with particles included?
Yes. The Strep-Gold universal kit includes streptavidin gold on the strip, and the fluorescent universal kit offers a fluorescent readout.
Can Attogene help develop a custom assay?
Yes. Our lateral flow development services can build a custom test around your reporter. You can also prototype with blank 4.5 mm lateral flow strips.