, , ,

Caffeine Rapid Detection Kit (Urine Compatible)

$149.00

  • Screening of Caffeine in liquid samples at >10 ppb
  • Format: 10 tests (5 tests/5 controls)
  • Sample Dilution Buffer
  • Sample Running Buffer
  • Negative Control
149149
Sample Dilution Buffer
Optional extra sample buffer for this rapid test kit. Your kit already includes enough buffer for its tests; add a 5 or 10 pack if you need to dilute samples that read above the kit's range, re-run samples, or keep spare buffer on hand for field work.
$60.00
$100.00
Addons Price: $0.00
Total Price: $149.00
- +

The Attogene Caffeine Rapid Detection Kit (AU2067, Urine Compatible) is a caffeine rapid test that screens urine, water and drinks for caffeine in 15 minutes. It is a single-use lateral flow cassette test with a cutoff of 10 ppb caffeine in the buffer tube after the sample is added. Results are read by eye or with a lateral flow reader, always against a negative control run at the same time. Each kit contains 10 cassettes for 5 samples, each with its own negative control.

The sample volume, from 1 to 20 µL, sets the detection range: about 110 ppb in the sample with 20 µL of urine or water, and about 2 ppm with 1 µL of a drink.

Key Features

  • 15-minute result: read at 15 minutes, and no later than 20 minutes.
  • 10 ppb cutoff: caffeine at or above 10 ppb in the buffer tube is detected.
  • Adjustable range: the sample volume (1–20 µL) sets the concentration in the sample that matches the cutoff.
  • Multiple sample types: urine, water, coffee, tea, soft drinks, energy drinks and other drinks.
  • Paired negative control: every sample is read side by side with its own negative control cassette.
  • Room-temperature storage: all components are stored at room temperature.

What’s in the Kit

Each caffeine rapid test kit includes:

  • 10 caffeine cassettes, each in a foil pouch with desiccant
  • 10 Sample Dilution/Running Buffer tubes, 200 µL each (one per sample is used as the negative control)
  • 10X Sample Dilution Buffer, 10 mL, for diluting concentrated samples
  • 10 fixed-volume pipettes, 200 µL
  • Manual

You will also need: a timer, a permanent marker, disposable gloves for urine samples, a 1–20 µL pipettor and tips, and clean sample cups. To dilute concentrated samples you will also need clean tubes, a 20–200 µL pipettor and ultrapure or distilled water. A caffeine standard of known concentration is optional, as a positive control or to estimate concentration. Extra Sample Dilution Buffer is available in packs of 5 or 10.

Who It’s For

  • Research laboratories screening urine or water samples for caffeine
  • Beverage and food labs checking drinks for caffeine, including decaffeinated products
  • Teams that want a fast screen before confirming by HPLC or LC-MS/MS
  • Assay developers who need numerical T/C readings on a lateral flow reader

How to Use the Caffeine Rapid Test

  1. Bring cassettes, buffers and samples to room temperature. Lay two cassettes flat for each sample and label them, for example “NC” and the sample name.
  2. Negative control: with a new 200 µL pipette, add buffer from a Sample Dilution/Running Buffer tube to the sample port of the NC cassette. Start the timer.
  3. Sample: add 20 µL of urine or water, or 1 µL of a drink, to a new buffer tube. Mix 3 times with a new 200 µL pipette.
  4. Add the mix to the sample port of the sample cassette, all at once.
  5. Leave both cassettes flat and undisturbed, and read them side by side at 15 minutes.

Let hot drinks cool and stir carbonated drinks until they stop fizzing. For samples above the range, dilute them in 1X Sample Dilution Buffer (1 part 10X buffer plus 9 parts distilled water), test again and multiply by the dilution factor. See the AU2067 instruction manual (PDF) for full details.

Suggested Procedure for Tea and Brewed Drinks

  1. Brew normally, or to a fixed recipe. Testing the drink as it would actually be consumed answers the real question: “is this cup decaf?” For results that are comparable between samples, use a fixed ratio, for example 2 g of tea in 200 mL of hot water for 3–5 minutes.
  2. Cool to room temperature before testing. Hot samples can affect antibody binding and how the sample flows along the strip.
  3. Remove polyphenols (optional, but likely needed for green tea). Catechins such as EGCG bind proteins and are the main interference risk in a lateral flow test. They can cause false positives or weak, smeared lines, especially in a competitive format like this one. A simple field fix is a pre-weighed scoop of PVPP (not included): swirl for a few minutes, then let it settle or pass the sample through a filter tip.
  4. Mix with the running buffer and apply as the instructions say.

Reading Results

The test is a competitive immunoassay. Gold-labeled anti-caffeine antibody binds a caffeine-BSA test line (T) when no caffeine is present. Caffeine in the sample blocks the antibody, so the test line gets weaker and the control line (C) gets darker. Unlike most lateral flow tests, a faint or missing C line is normal for a negative sample in this kit.

  • Negative: looks like the negative control; strong T line, C line faint or absent.
  • Low positive: T still darker than C, but C clearly darker than on the negative control (10 ppb to about 100 ppb in the tube).
  • Positive: C as dark as or darker than T, or no T line (about 100 ppb or more in the tube).
  • Invalid: no line on a cassette, or no strong T line on the negative control.

On a lateral flow reader, a sample T/C less than half the negative control T/C indicates caffeine at or above the cutoff.

Frequently Asked Questions

What samples can the caffeine rapid test screen?

Urine, water, coffee, tea, soft drinks, energy drinks and other drinks. Drinks with milk, cream, syrups or pulp have not been evaluated.

How much sample do I add?

Use 20 µL for urine and water, and 1 µL for drinks. Intermediate volumes of 10, 5 or 2 µL shift the cutoff in the sample to about 210 ppb, 410 ppb or 1 ppm.

Will decaffeinated coffee read positive?

Often, yes. Decaffeination removes most but not all caffeine, and decaffeinated coffees usually contain more than 2 ppm, so they typically read low positive or positive with 1 µL of drink.

Does the test cross-react with other compounds?

Cross-reactivity with theobromine, theophylline and paraxanthine has not been evaluated, and these may contribute to a positive result.

Can this caffeine rapid test be used for diagnostic or forensic testing?

No. It is a research-use screening test and is not for diagnostic or forensic use. Confirm results used for any decision by HPLC, LC-MS/MS or another conventional method.

Is there a version for testing drinks in the field?

Yes. The Caffeine Rapid Detection Kit – Field Use (AU2067-01) tests drinks without a pipettor, using fixed-volume pipettes and a 1:400 dilution. For nicotine screening, see the Nicotine Rapid Detection Kit (AU2068).

For research use only. Not for use in diagnostic procedures. Questions? Call 512-333-1330 or email sales@attogene.com.

tests

5 tests, 20 tests

You may also like…

Shopping Cart