Capsaicin Detection Kit (Rapid – Lab)

$149.00

  • Rapid dipstick test for detecting Capsaicin in liquid samples, crushed peppers, pepper extracts
  • Detects 10 ppb and greater Capsaicin in the sample
  • Can be used to convert to Scoville heat units
- +

The Capsaicin Detection Kit, Rapid-Lab (AU2046) is a capsaicin lateral flow test for screening liquid hot sauce, pepper paste, pepper extracts and other liquid samples in the lab. It detects capsaicin at 10 to 20 ppb (final concentration in the assay buffer) with a 15-minute run time, and results can be converted to Scoville heat units (SHU). Each kit contains 10 cassettes.

Capsaicin is the main compound that gives chili peppers their heat. The Scoville scale runs from 0 SHU for a bell pepper to about 16 million SHU for pure capsaicin. The original Scoville method relies on human taste panels, and instrument methods such as HPLC and GC-MS are precise but costly and slow. Attogene developed this test so hot sauce makers and labs can measure capsaicin with minimal equipment or training.

Key Features

  • Sensitive: Detects capsaicin at 10 to 20 ppb in the assay buffer.
  • 15-minute run time: Fast results with minimal equipment.
  • Little sample preparation: 1 to 15 µL of sample goes straight into the running buffer.
  • Wide working range: Dilute strong samples in the supplied 1X PBS to bring them into range.
  • Semi-quantitative option: Run capsaicin standards alongside samples, or use a lateral flow reader such as the RDS-2500 from Detekt Biomedical for numerical results.
  • SHU conversion: Capsaicin concentration can be converted to Scoville heat units with a simple factor.

What’s in the Kit

Each capsaicin lateral flow test kit includes:

  • 10 capsaicin test cassettes, individually foil-pouched with desiccant
  • 10 tubes of sample dilution/running buffer, 150 µL each
  • 10 fixed volume pipettes, 75 µL
  • 10 mL 1X dilution buffer

You will also need: a pipette that can dispense 1 to 10 µL, a timer, a marker for labeling and a capsaicin control or standard.

Who It’s For

  • Hot sauce and food manufacturers, from hobby to industrial scale, checking capsaicin levels during or after production.
  • Food QA and contract labs that want a fast alternative to HPLC or GC-MS screening.
  • Researchers working with pepper extracts and capsaicin-containing products.

How to Use the Capsaicin Lateral Flow Test

  1. Wearing gloves, remove cassettes from their foil pouches and label them.
  2. Negative control first: transfer 150 µL of sample dilution/running buffer into the sample port of a cassette using the 75 µL pipette twice. Start a 15-minute timer.
  3. Sample: pipette 1 to 15 µL of sample into a tube of sample dilution/running buffer and mix gently up and down 3 times with the 75 µL pipette.
  4. Transfer 150 µL from the tube into the sample port of a new cassette (two pipette draws). Start a 15-minute timer.
  5. Read the test and control lines by eye or with a lateral flow reader.

If a sample contains 20 ppm or more capsaicin, dilute it 2- to 10,000-fold or more in the 1X dilution buffer before testing. Full instructions are in the AU2046 manual (PDF).

Reading the Results

This is a competitive assay: the more capsaicin in the sample, the lighter the test line. Compare each sample cassette with the negative control cassette. A visible control line is required for a valid result.

  • Test line darker than or equal to the negative control: below the detection limit.
  • Test line lighter than the negative control: 10 to 20 ppb or greater.
  • No test line, control line only: 20 ppb or greater.

Read results within 5 minutes after the run is complete. For semi-quantitative results from 0 to 20 ppb, run capsaicin control solutions of known concentration at the same time as your samples.

From Capsaicin Concentration to Scoville Heat Units

Because pure capsaicin rates about 16 million SHU, each 1 ppm of capsaicin corresponds to roughly 15 to 16 SHU. Once you know the capsaicin concentration in the original sample, after correcting for any dilution, multiply by this factor to estimate heat. In Attogene’s technical article (PDF), the test detected at least 10 ppb capsaicin with a reader and distinguished a 450 SHU sauce from a 7 million SHU sauce.

Frequently Asked Questions

What samples work with the capsaicin lateral flow test?

The kit is intended for liquid hot sauce, pepper paste, pepper extracts and other liquid samples.

How sensitive is it?

It detects capsaicin at 10 to 20 ppb, measured as the final concentration in the assay buffer.

Does the kit give an SHU value directly?

No. The capsaicin lateral flow test measures capsaicin; you convert that concentration to SHU using the factor above. Running capsaicin standards or using a reader improves the estimate.

What is the difference between the Rapid-Lab and Rapid-On-site kits?

This lab kit uses small sample volumes and requires a 1 to 10 µL pipette. The on-site capsaicin test kit (AU2046-01) includes a pre-measured 100X dilution system and transfer tips, so no lab pipette is needed.

How should the kit be stored?

Store at 2 to 30 °C and keep cassettes sealed until use.

See our full range of lateral flow products.