The QDot Staphylococcal Enterotoxin B Lateral Flow Test Kit (AU2048-01) is an SEB fluorescent lateral flow test for screening milk, meat rinsate and water for staphylococcal enterotoxin B. It uses an anti-SEB antibody bound to streptavidin quantum dots (QDots) developed by Attogene, and with reader-assisted analysis it detects SEB down to 20 ppb in milk, 10 ppb in meat rinsate and 1 ppb in water. Each kit contains 10 cassettes, and results are ready in 10 to 15 minutes.
SEB is a heat-stable toxin produced by Staphylococcus aureus and a common cause of staphylococcal food poisoning. Because the toxin can survive cooking and pasteurization, testing the food or water itself is often more informative than testing for the bacteria alone.
Key Features
- Quantum dot labels: Fluorescent QDot conjugates give higher sensitivity than the colloidal gold version of this test.
- Multiple sample types: Sample preparation instructions for milk, chicken or meat rinsate and water.
- Low detection levels: 20 ppb in milk, 10 ppb in meat rinsate and 1 ppb in water with a reader.
- Fast: 10 to 15 minute incubation after the sample is applied.
- Flexible readout: Read by eye, by photograph or in a fluorescent lateral flow reader.
- Room-temperature storage: Store the kit at 2 to 30 °C.
- Development-friendly: Useful as a starting point for your own SEB lateral flow assay development.
What’s in the Kit
Each SEB fluorescent lateral flow test kit includes:
- 10 lateral flow cassettes
- 5 tubes of sample diluent buffer
- 5 tubes of negative control
- 5 fixed volume pipettes
- Instruction manual
You will also need: a centrifuge (for clarifying milk), stomacher bags or wide-mouth bottles (for rinsate collection), sample bottles for storage, swabs for surface sampling and a scale for weighing tissue. A fluorescent lateral flow reader is optional; ask us about a reader option.
Who It’s For
- Food safety and QA labs screening dairy and meat products for SEB.
- Water testing labs that need a low ppb screen for SEB.
- Researchers and assay developers building or benchmarking fluorescent lateral flow tests for staphylococcal enterotoxins.
How to Use the SEB Fluorescent Lateral Flow Test
Sample preparation
- Milk: Run as is, or for higher sensitivity freeze/thaw the milk and centrifuge at 10,000 RCF for 15 minutes. Untreated milk will not reach 20 ppb detection.
- Chicken and meat: Place about 25 g of tissue in a stomacher bag or wide-mouth bottle, add 100 mL of sterile distilled water and shake vigorously. Use the rinsate within a few hours or freeze at -20 °C.
- Water: Test directly.
Running the test
- Using a fixed volume pipette, dispense the full contents of a negative control tube onto the sample port of a cassette.
- Add 200 µL of sample to a tube of sample diluent buffer and mix thoroughly (1:1 dilution).
- Dispense 200 µL of the mixed sample into the sample port of a new cassette.
- Incubate for 10 to 15 minutes.
- Read the cassette by eye, by photograph or in a lateral flow reader.
Full instructions are in the AU2048-01 manual (PDF).
Reading the Results
This is a sandwich-style test: the more SEB in the sample, the brighter the test line compared with the negative control cassette. The control line should stay visible and its intensity does not drop as the test line increases.
Very high toxin levels can cause a hook effect. If you expect SEB but see no test line, run a serial dilution of the sample. Read cassettes at the specified time, as line intensities can change as strips dry.
How It Works
Quantum dots are bright, stable fluorescent labels. In a study published by Attogene and Nanohmics scientists in the Journal of Fluorescence, QDot labels improved SEB lateral flow sensitivity at least four-fold over a colloidal gold version of the same test, and the method addressed QD clumping and signal loss in chicken rinsate and whole milk. Read the study (PDF).
Frequently Asked Questions
How sensitive is the SEB fluorescent lateral flow test?
With reader-assisted analysis, the kit detects SEB down to 20 ppb in milk, 10 ppb in meat rinsate and 1 ppb in water.
Do I need a fluorescent reader?
No. Cassettes can be read by eye or photographed, but a fluorescent lateral flow reader gives objective readings and the best sensitivity. Contact us about a reader option.
How is this different from the colloidal gold SEB kit?
The colloidal gold SEB lateral flow kit (AU2048) is a lower-cost visual test for water. The SEB fluorescent lateral flow test adds milk and meat rinsate and offers higher sensitivity.
How long does a test take?
Incubation is 10 to 15 minutes after the sample is applied. Milk centrifugation adds about 15 minutes.
How should the kit be stored?
Store at 2 to 30 °C and keep cassettes sealed in their foil pouches until use.
Browse more rapid tests on our lateral flow products page.
For research use only. Not for use in diagnostic procedures. Questions? Call 512-333-1330 or email sales@attogene.com.